| J Dent Anesth Pain Med. 2020 Apr;20(2):73-81. English. Published online Apr 27, 2020. https://doi.org/10.17245/jdapm.2020.20.2.73 | |
| Copyright © 2020 Journal of Dental Anesthesia and Pain Medicine | |
|
Bong Jin Kang | |
|
1Department of Anesthesiology and Pain Medicine, College of Medicine, Dankook University, Cheonan, Korea. | |
|
2Department of Anesthesiology and Pain Medicine, Asan Medical Center, University of Ulsan College of Medicine, Seoul, Korea. | |
|
3Department of Statistics, Ewha Womans University, Seoul, Korea. | |
|
4Department of Clinical Pharmacology and Therapeutics, Asan Medical Center, University of Ulsan College of Medicine, Seoul, Korea. | |
| Received March 13, 2020; Revised April 07, 2020; Accepted April 09, 2020. | |
|
This is an Open Access article distributed under the terms of the Creative Commons Attribution Non-Commercial License (http://creativecommons.org/licenses/by- | |
|
Abstract
| |
|
Background
The international organization for standardization (ISO) 80601-2-61 dictates that the accuracy of a pulse oximeter should be assessed by a controlled desaturation study. We aimed to characterize the relationship between the fraction of inspired oxygen (FiO2) and peripheral oxygen saturation (SpO2) using a turnover model by retrospectively analyzing the data obtained from previous controlled desaturation studies.
Materials and Methods
Each volunteer was placed in a semi-Fowler's position and connected to a breathing circuit to administer the hypoxic gas mixture containing medical air, oxygen, nitrogen, and carbon dioxide. Volunteers were exposed to various levels of induced hypoxia over 70–100% arterial oxygen saturation (SaO2). The study period consisted of two rounds of hypoxia and the volunteers were maintained in room air between each round. FiO2 and SpO2 were recorded continuously during the study period. A population pharmacodynamic analysis was performed with the NONMEM VII level 4 (ICON Development Solutions, Ellicott City, MD, USA).
Results
In total, 2899 SpO2 data points obtained from 20 volunteers were used to determine the pharmacodynamic characteristics. The pharmacodynamic parameters were as follows: kout = 0.942 1/min, Imax = 0.802, IC50 = 85.3%, γ = 27.3.
Conclusion
The changes in SpO2 due to decreases in FiO2 well explained by the turnover model with inhibitory function as a sigmoidal model. |
|
Keywords:
Mathematical Computing; Oximetry; Pharmacology
|
|
|
INTRODUCTION
|
A pulse oximeter noninvasively measures arterial blood oxygen saturation in dental care area. Respiratory management of patients is carried out by referring to the value of peripheral oxygen saturation (SpO2) measured by the pulse oximeter [1, 2]. Therefore, it is important to ensure that pulse oximeters have high accuracy. The accuracy criteria for the pulse oximeter equipment are provided in the International Standard ISO 80601-2-61, and the U.S. Food and Drug Administration (FDA) only permits pulse oximeters that meet these criteria [3, 4]. According to the ISO 80601-2-61, the SpO2 accuracy of a pulse oximeter should be assessed through a controlled desaturation study by comparing to the gold standard measurements of blood arterial oxygen saturation (SaO2) obtained by a CO-oximeter [3]. In controlled desaturation studies, the range of SaO2 is typically 70–100%, and the subjects are temporarily exposed to a fraction of inspired oxygen (FiO2) lower than that in room air. In such situations, SpO2 seems to decrease as FiO2 decreases. Quantifying the relationship between FiO2 and SpO2 allows us to better explain the two relationships. In general, a turnover model appropriately describes the relationship between drug concentration and response [5]; particularly, the model is able to explain the concentration-response relationship when an endogenous biomarker is the response [5, 6].
The aim of this study was to characterize the relationship between FiO2 and SpO2 using a turnover model by retrospectively analyzing the data obtained from a controlled desaturation study.
|
METHODS
|
In order to evaluate the quantitative relationship between FiO2 and SpO2, we utilized data obtained from a previous controlled desaturation study. Briefly, the protocol of the controlled desaturation study is as follows.
1. Protocol of the controlled desaturation study
The study had a single-center, non-randomized design. The protocol was approved by the Asan Medical Center Institutional Review Board (approval number: 1st study: 2014-1194, 2nd study: 2016-0069). Written informed consent was obtained from all volunteers, whose inclusion criteria included aged 20–50 yr, carboxyhemoglobin < 3%, methemoglobin < 2%, and total hemoglobin concentration > 10 g/dl. Volunteer exclusion criteria included known history of respiratory or cardiovascular disease, smoking habits, evidence of pregnancy, history of syncope, diabetes mellitus, and body mass index ≥ 35 kg/m2. The volunteers were fully informed of the study protocols and completed health assessment questionnaires prior to enrollment.
In an operating room equipped for a controlled desaturation study, the volunteers were monitored using electrocardiography, end-tidal carbon dioxide partial pressure, FiO2, and non-invasive blood pressure by using Carescape® B850 (GE Healthcare, Milwaukee, WI, USA). Throughout the study period, all aforementioned data were continuously downloaded to personal computers. Each volunteer was placed in a semi-Fowler's position and connected to a breathing circuit to administer the hypoxic gas mixture containing medical air, oxygen, nitrogen, and carbon dioxide. For frequent blood sampling, an arterial cannula was placed in a radial artery of each volunteer, and a reusable finger probe of a pulse oximeter (OxiMax® N-600x, Medtronic, Boulder, CO, USA) was placed on a finger on the cannulated side. In order to prevent hypoperfusion, an air warmer (Bair Hugger™, 3M™, St. Paul, MN, USA) was applied to the hand with the finger probe. Each volunteer was exposed to various levels of induced hypoxia over 70–100% SaO2. Each plateau of oxygen saturation was maintained for at least 30 s until stabilization, after which 1 ml of arterial blood was drawn into a heparinized syringe. The study period consisted of two rounds of hypoxia and the volunteers were maintained in room air between each round.
2. Data selection of FiO2 and SpO2
The individual data files of FiO2 and SpO2 were recorded continuously during the study period. The FiO2 and SpO2 values were updated every 10 s and 1 s, respectively. For pharmacodynamic analysis, data points from every 20 s were selected.
3. Population pharmacodynamic analysis
A population pharmacodynamic analysis was performed with the NONMEM VII level 4 (ICON Development Solutions, Ellicott City, MD, USA). A turnover model using the ADVAN 13 subroutines and first-order conditional estimation with interaction was fitted to FiO2-SpO2 pair data. As the turnover model is mainly used to explain the relationship of increasing or decreasing response with increasing concentration, the processed FiO2 was calculated for this application. Processed FiO2 was defined as 100 minus FiO2.
where kin is the turnover rate constant, kout is the fractional turnover rate constant, and SpO2_baseline is the baseline SpO2. I(c) is an inhibitory function to explain the relationship between the processed FiO2 and SpO2. I(c) was expressed as a linear or a sigmodal model as follows,
where slope is the slope factor in the inhibitory effect of FiO2 on the production of the response (SpO2), Imax is the maximum fractional inhibitory ability of processed FiO2 to affect SpO2,IC50 is the processed FiO2 producing 50% of Imax, γ is the steepness of the processed FiO2 versus SpO2 relationship. To reduce the number of parameters to be estimated, SpO2_baseline was set to 100. Inter-individual random variabilities of pharmacodynamic parameters were estimated by assuming a log-normal distribution. Diagonal matrices were estimated for the various distributions of η, where η represents inter-individual random variability with a mean of zero and variance of ω2. Additive residual error model was evaluated during the model building process. NONMEM computed the minimum objective function value (OFV), a statistical equivalent to the −2 log-likelihood of the model. An α level of 0.05, which corresponds to a reduction of 3.84 in the OFV (chi-square distribution, degree of freedom = 1, P < 0.05), was used to distinguish between hierarchical models [7]. The covariates analyzed were age, sex (1: male; 0: female), and race (1: Asian, 0: African). Non-parametric bootstrap analysis was used for internal validation of the models (fit4NM 3.3.3, Eun-Kyung Lee and Gyu-Jeong Noh; http://cran.r-
4. Statistics
Statistical analyses were conducted using R (version 3.5.2; R Foundation for Statistical Computing, Vienna, Austria) or SigmaStat version 3.5 for Windows (Systat Software, Inc, Chicago, IL, USA). Data are expressed as mean ± standard deviation for normally distributed continuous variables, median (25–75%) for non-normally distributed continuous variables, or counts for categorical variables.
|
RESULTS
|
Of the 24 volunteers enrolled, four were excluded because of a technical error in FiO2 data file storage. Hence, 20 volunteers were included in the pharmacodynamic analysis. The characteristics of these volunteers are summarized in Table 1. In total, 2899 SpO2 data points from 20 volunteers were used to determine the pharmacodynamic characteristics. Time courses of the processed FiO2 and SpO2 are shown in Figure 1. As the FiO2 decreased, the SpO2 values tended to decrease as well. The SpO2 values were maintained between 65–100% in each round in which hypoxic gas was supplied.
|
|
The turnover model well described the time course of SpO2 (Appendix). In particular, expressing the inhibitory function as a sigmoid model rather than a linear model further reduced the objective function value (OFV: 14495.97 for the linear model, 13179.24 for the sigmoidal model). No significant covariates for the pharmacodynamic parameters were observed. Figure 2 shows the plots of the predicted versus observed SpO2 in the volunteers with the lowest or highest absolute values of the individual mean of weighted residuals. Table 2 shows the population pharmacodynamic parameter estimates, inter-individual variability, and median parameter values (2.5–97.5%) of the non-parametric bootstrap replicates of the final pharmacodynamic model of SpO2. Predictive checks of the final pharmacodynamic model are presented in Figure 3. In total, 19.5% of the data were distributed outside of the 90% prediction intervals of the predictive check.
|
|
|
|
DISCUSSION
|
The changes in SpO2 due to decrease in FiO2 under room air was well explained by the turnover model with inhibitory function as a sigmoidal model. In the setting of a controlled desaturation study, the potency of FiO2 required to reduce SpO2 from 100% to 70% was 14.7%.
The pulse oximeter is an essential monitoring medical device across many fields of medicine [8]. The SpO2 accuracy of a pulse oximeter equipment should be a root-mean-square difference of less than or equal to 4.0% SpO2 over the range of 70%–100% SaO2 [3],
where Arms describes the combined bias and precision of SpO2 readings, SpO2ij means the jth measured SpO2 value of the ith volunteer, and SRij refers to the jth measured standard reference value of the ith volunteer. The standard reference value is determined by SaO2, which is measured by using a CO-oximeter for arterial blood collected at the same time when SpO2 is observed. The common manufacturing literature claim for Arms for pulse oximeters is ± 2–3% over the range of 70–100% SpO2 [8]. The total number of acceptable SpO2-SaO2 pairs obtained during clinical trials should be sufficient to statistically validate the specified SpO2 accuracy. Typically, at least 10 volunteers are recruited, and at least 20 arterial blood samples per volunteer are obtained and analyzed with at least 200 data pairs. In addition, the distribution of the SaO2 values should have similar density over the entire required range; for example, the ranges of 70–79%, 80–89%, and 90–100% SaO2 should each have approximately 1/3 of the total data. Particularly, the complexion of the study participants should be specified because the accuracy of SpO2 depends on the complexion, with dark skin pigmentation resulting in an overestimation of arterial oxygen saturation especially at low saturation in some pulse oximeters [9]. With the Nellcor N-595 device (Medtronic, Boulder, CO, USA), there was a significant difference in bias, defined as SpO2 minus SaO2, between light skin and dark skin in the 60–100% SpO2 range [9]. For this reason, the U.S. FDA recommended that controlled desaturation studies should have subjects across a range of skin pigmentations, including at least 2 darkly pigmented subjects or 15% of the subject pool, whichever is larger [4]. However, in Korea, such desaturation study is not required to permit the pulse oximeter equipment.
If equilibrium is quickly established between the concentration of plasma and the response, then the pharmacological effect is immediately apparent; in such case, direct pharmacodynamic models such as the linear model or sigmoid Emax model can be applied. However, several drug responses can be considered as indirect in nature [10]. The turnover model can be appropriate for use when there is a delay between concentration and response [5, 10]; accordingly, the relationship between end-tidal carbon dioxide and regional cerebral oxygen saturation with delay was also quantified by the turnover model [6]. The net baseline effect is the balance between the apparent rate of “production” of the response and the rate of “removal” of the response [5]. In situations where the production and removal of response are complex, one step is rate-limiting, and these are represented by first-order rate constants kin and kout, respectively [5]. The turnover model is divided into four basic models depending on whether the response increases or decreases with increasing concentration [10]. The structural formulas of the four models are as follows [10].
Model 1: inhibition of production
Model 2: inhibition of loss
Model 3: stimulation of production
Model 4: stimulation of loss
, where kin is the turnover rate constant, kout is the fractional turnover rate constant, and Imax and Emax are maximal inhibitory and stimulatory effects attributed to drugs, respectively. IC50 and EC50 are drug concentration producing 50% of maximum stimulation and inhibition at effect site, respectively. R is a response variable and γ is the sigmoidicity factor. In the current study, basic model 1 was applied. Also, the sigmoidal model was more suitable than the linear model for use as an inhibition function. Graphically, the predicted value estimated by the sigmoidal model was also closer to the observed value. Physiologically, it is difficult to interpret a linear decrease in SpO2 as FiO2 decreases. The oxygen–hemoglobin dissociation curve that plots the proportion of hemoglobin in its saturated form on the vertical axis against the prevailing oxygen tension on the horizontal axis has a sigmoid shape; therefore, it is also desirable to apply the sigmoidal model for inhibitory function.
The following issues should be considered as limitations of this study. First, the observed FiO2 may be somewhat inaccurate as intubation was not used and the volunteers breathed the mixed hypoxic gas through their mouth while their noses were blocked with a clothespin. Therefore, the volunteers' exhalation may have affected the FiO2. However, a desaturation study involving volunteers does not generally require endotracheal intubation; also, as shown in Figure 1, the changes in FiO2 during the study period were generally acceptable. Second, interindividual variability (IIV) in pharmacodynamic parameters could not be explained by covariates. As depicted in Table 2,IC50 had approximately 45% IIV. However, in volunteer-based studies, the covariates are often not included because the demographic characteristics are homogenous. Covariates may explain the IIV in pharmacokinetic and/or pharmacodynamic parameters in studies involving patients with varying characteristics [6, 11]. Importantly, the number of volunteers in this study was too small to include covariates.
In conclusion, by using the data obtained in previous controlled desaturation studies, we observed that the SpO2 value tended to decrease as the FiO2 decreased. The relationship between FiO2 and SpO2 was well described by the turnover model with inhibitory function as a sigmoidal model. The potency of FiO2 required to reduce SpO2 from 100% to 70% was 14.7%.
|
Notes
|
AUTHOR CONTRIBUTIONS:
Bong Jin Kang: Conceptualization, Formal analysis, Investigation, Methodology, Supervision, Visualization, Writing — original draft.
Myojung Kim: Data curation, Methodology, Visualization, Writing — original draft, Writing — review & editing.
Ji-Yeon Bang: Conceptualization, Methodology, Project administration, Resources, Software, Supervision, Validation, Visualization, Writing — original draft, Writing — review & editing.
Eun-Kyung Lee: Methodology, Formal analysis, Writing — review & editing.
Byung-Moon Choi: Conceptualization, Data curation, Formal analysis, Methodology, Investigation, Resources, Software, Supervision, Validation, Visualization, Writing — original draft, Writing — review & editing.
Gyu-Jeong Noh: Conceptualization, Data curation, Formal analysis, Funding acquisition, Methodology, Resources, Software, Supervision, Validation, Visualization, Writing — original draft, Writing — review & editing.
FUNDING:This study was supported by a grant (MFDS2015-20766) from the Ministry of Food and Drug Safety, Cheongju-si, Republic of Korea.
DECLARATION OF INTEREST:None.